File name RELEASE_NOTE.2.5.1351.6139 Release Note for 2.5.1351.6139.txtB2900A 2.5.1351.6139 RELEASE NOTE
- NEW FEATURES
Scripting (License is needed.)
LED Demonstration Program (Press [More]-[LED Demo])
- FIXED DEFECTS
#162299 Window2 text data is hidden when local key was pressed.
#184063 I/O hangs when transferring large amount of data.
#187575 Intermittent hang-up when sweep measurement is aborted by output protection.
#193236 Unable to abort infinite waveform output when both two channels are busy.
B2900A 2.0.1312.2910 RELEASE NOTE
- FIXED DEFECTS
#96216 Ib-Ic curve capability in Graph View Mode.
#96222 Request to support "Infinity" as Trigger Count.
#102948 Unexpected program memory status is shown.
#149042 Setting conflict error doesn't happen when running on the front panel.
#149362 Unexpected jumps at joint of the curves when repeating curves on list sweep with ARM:INFINITY.
#149694 Unexpected hang-up after measurement with long wait and time constant.
#158151 Cannot launch the web instrument on IPv6.
#158884 The instrument shows "Fan speed is too fast" in RACK mode.
#162283 Unexpected pop-up window is shown if USB-A (front) is connected to USB-B (rear).
#177173 Application Error occurs during Sweep Measurements
#179938 Hangs up by front panel operation during SCPI command is in progress.
#179941 Unexpected trace configuration is shown on Trace Buffer Setup dialog.
#182858 Unexpected query result on :SENSe:DATA? command.
B2900A 2.0.1225.1717 RELEASE NOTE
- NEW FEATURES
LXI 1.4 (LXI Core, LXI HiSLIP, and LXI IPv6)
- FIXED DEFECTS
#77088 Wants to save trigger mode.
#77090 Wants to specify the increment of knob rotating.
#86917 Crash on repetitive list sweep.
#87578 Wants to input "10pA" not "0.01nA".
#88223 Hang up when "ABORt" command is sent during being initiated.
#90979 Unexpected waveform preview on List Sweep/Edit dialogs.
#95675 :ARM:TRAN:COUN? INF does not cause an error.
#95791 Unexpected self-test failure on low temperature.
#97269 Slow program memory execution time.
#99534 LAN setting get corrupt when cable is unplugged in static IP mode.
#100315 Unexpected source level when changing source mode.
#100848 Unexpected glitches when current range is changed to 1A from 100mA.
#102233 Source mode is not restored if it was canceled.
#103657 Incorrect SCPI error messages are shown on the front panel.
#104236 Empty file is created even if saving configuration failed on the front panel.
B2900A 1.0.1144.9870 RELEASE NOTE
- FIXED DEFECTS
#4650 When mDNS is disabled with no LAN connection, AgilentLxiWebService.exe dies.
#63772 Incorrect source values are fetched when ARM count is INFinite.
#64217 Incorrect error message for "IADC Lost" event.
#70493 Unexpected thin-out result on Graph/Roll view.
#74773 Cannot show CH2 data captured during remote state with remote display off
#74786 Unexpected voltage/current output when stepping voltage/current list by EXTn signals.
#76607 Does not output trigger when sp |